A calibration study of therapeutic ultrasound units. (Technical Report).


Therapeutic ultrasound Therapeutic ultrasound is a technique that uses high-frequency sound waves (ultrasound) to speed healing in injured joint or muscle tissue. The frequency used is typically 1-3 Mhz.  (US) is a frequently used modality modality /mo·dal·i·ty/ (mo-dal´i-te)
1. a method of application of, or the employment of, any therapeutic agent, especially a physical agent.

2.
 in the practice of physical therapy. (1,2) Much research has been conducted on the effects of US on living tissue, and both positive and negative effects have been reported. These effects are dependent on the intensity and frequency of the US applied to the tissue. For example, an increase in intracellular intracellular /in·tra·cel·lu·lar/ (-sel´u-ler) within a cell or cells.

in·tra·cel·lu·lar
adj.
Occurring or situated within a cell or cells.
 calcium, which leads to faster tissue healing, was shown to occur in human tissue in cultured medium at intensities of 0.5 to 0.75 W/[cm.sup.2] with a pulsed frequency of 20%. (3) Other beneficial effects such as increased soft tissue extensibility, (4,5) decreased pain levels, (6,7) and faster and stronger repair of tendon tendon, tough cord composed of closely packed white fibers of connective tissue that serves to attach muscles to internal structures such as bones or other muscles.  injuries (8-11) have also been reported.

Not all effects of US on tissue are beneficial. Ultrasound delivered at 1 W/[cm.sup.2] (energy in watts divided by the size of the soundhead's effective radiation area in centimeters squared) and a 1-MHz frequency has been shown to increase the temperature in the human gastrocnemius muscle gastrocnemius muscle

see Table 13.


gastrocnemius muscle rupture, gastrocnemius muscle avulsion
the muscle may have torn away from its insertion, in which case the tendon will be slack, or it may be a complete or partial separation
 by 0.2 [degrees] C per minute? Although the rise in tissue temperature can be beneficial, such as for pain reduction, addition of excessive energy to the tissue poses potential risks. For example, bone damage, including inhibition of bone growth (13) and damage to bone marrow (14) has been shown to occur in dogs at intensities above 3 W/[cm.sup.2].

Due to its potential beneficial and deleterious deleterious adj. harmful.  effects, which depend on the intensity and duration of application, the US soundhead output should be what is expected. The effects of US on tissue are intensity-dependent; too low a dose will have no clinical effect, whereas too high a dose can be damaging. (15) Ultrasound used inappropriately may be at best ineffective or at worst damaging to the patient. (16)

In order to help ensure that US equipment is properly calibrated cal·i·brate  
tr.v. cal·i·brat·ed, cal·i·brat·ing, cal·i·brates
1. To check, adjust, or determine by comparison with a standard (the graduations of a quantitative measuring instrument):
, several operational standards for US machines' outputs have been published. The US Department of Health, Education, and Welfare (17) specifies that US output should be within [+ or -] 20% of the intensity indicated on the US machine. The Canadian government (18) and the International Electrotechnical Commission See IEC.

(standard, body) International Electrotechnical Commission - (IEC) A standardisation body at the same level as ISO.
 (19) have set calibration standards of [+ or -] 30%. All 3 standards specify that timers be accurate to within [+ or -] 10%.

Researchers have reported that a large number of therapeutic US machines used clinically were not within the standard. In 2 studies, (20,21) US calibration accuracy was assessed using the [+ or -] 20% standard. A study described in 1974 showed that 49 (85%) of the 58 US machines tested were not within this standard. (20) A 1981 study demonstrated that 21 (81%) of 26 machines tested were outside the standard. (21)

In the most recent studies, the less stringent [+ or -] 30% calibration standard was used. A report in 1987 showed that 24 (56%) of the 43 machines tested were outside of this calibration standard. (22) Another report published in 1992 showed that 59 (69%) of the 85 US machines tested were not within this standard. (23) A study published in 1997 showed that, out of 31 machines, "almost all" of the US machines tested were outside of this standard?

Results from previous research dating back to the early 70s indicate that a number of US machines had outputs that were outside the standard. The purpose of our study was to ascertain whether therapeutic US machines were within the [+ or -] 20% standard for power output and the [+ or -] 10% standard for timer accuracy.

Method

We evaluated 83 US machines used in clinical facilities in 10 counties of northwest Texas over a period of 3 months, from May until August 2000. The inclusion criteria
For Wikipedia's inclusion criteria, see: What Wikipedia is not.


Inclusion criteria are a set of conditions that must be met in order to participate in a clinical trial.
 required the machines to be in use in a rehabilitation rehabilitation: see physical therapy.  facility and to be considered by the staff as appropriate for patient treatment. Personnel in the facilities were unaware of the US variables to be evaluated.

A short letter was sent to potential participants explaining the study and inviting their participation. The clinics were later contacted via phone to confirm participation and schedule appointment times. Testing occurred at the participating rehabilitation facilities. Measurements of power output and timer accuracy were obtained from all US units and compared with the American standard of [+ or -] 20% for power output and [+ or -] 10% for timer accuracy.

Equipment

A Bio-Tek Digital Ultrasound Wattmeter Wattmeter

An instrument that measures electric power. See Electric power measurement

A variety of wattmeters are available to measure the power in ac circuits. They are generally classified by names descriptive of their operating principles.
 (model UW-2) * was used to test the power output (Fig. 1). Bio-Tek Instruments Inc calibrated the UW-2 the week prior to the start of testing and certified the device to be accurate for a period of 1 year under normal use. This device measures output through a linear variable differential transmitter (LVDT LVDT Linear Variable Differential Transformer
LVDT Linear Variable Displacement Transducer
LVDT Linear Variable Differential Transducer
LVDT Linear Voltage Differential Transformer
LVDT Low Voltage Differential Transceiver
LVDT Low Voltage Differential Transducer
). Deflection deflection /de·flec·tion/ (de-flek´shun) deviation or movement from a straight line or given course, such as from the baseline in electrocardiography.

de·flec·tion
n.
1.
 of a spring-and-cone assembly in the fluid is measured by the LVDT through means of a movable core. This deflection is a measurement of the amount of ultrasonic ultrasonic /ul·tra·son·ic/ (-son´ik) beyond the upper limit of perception by the human ear; relating to sound waves having a frequency of more than 20,000 Hz.

ul·tra·son·ic
adj.
1.
 energy applied to the cone. The UW-2 measures watts only, so the actual measured readings from the wattmeter were divided by each soundhead's effective radiation area, as specified by the manufacturer, in order to convert the readings to watts per square centimeter centimeter (sĕn`tĭmē'tər), abbr. cm, unit of length equal to 0.01 meter, the basic unit of length in the metric system. The centimeter is the unit of length in the cgs system. It is approximately equal to 0. . The UW-2 has a resolution of 0.1 W and an accuracy of [+ or -] 10%.

[FIGURE 1 OMITTED]

A digital stopwatch was used to test the accuracy of the timers. Prior to testing, this stopwatch was checked against the US atomic clock atomic clock, electric or electronic timekeeping device that is controlled by atomic or molecular oscillations. A timekeeping device must contain or be connected to some apparatus that oscillates at a uniform rate to control the rate of movement of its hands or the  to ensure its calibration.

Procedure

A pilot study was conducted prior to data collection to determine the number of trials needed for reliable measurements. Based on a test-retest protocol, a perfect correlation coefficient Correlation Coefficient

A measure that determines the degree to which two variable's movements are associated.

The correlation coefficient is calculated as:
 (r) of 1.0 was obtained, with zero variance for any repeated measurements. Therefore, based on those results, one measurement per intensity setting and timer interval was sufficient to obtain reproducible readings.

Clinicians filled out a brief questionnaire providing general information about each US unit tested. This information included each machine's brand name, age, most commonly used intensities, and how often the machine was used per week. In addition, the survey included how often the machine was checked and which variables were tested.

The transducer transducer, device that accepts an input of energy in one form and produces an output of energy in some other form, with a known, fixed relationship between the input and output.  well of the UW-2 was filled with 55 mL of degassed water for the coupling medium. A dissolved oxygen test kit was used to test each container of degassed water prior to use to ensure that the oxygen content was less than 2 ppm. More than 4 ppm of oxygen reduces the accuracy of the power output measurements by interfering with the transmission of the ultrasonic energy. (21)

Particular care was used to ensure that the transducer head was properly placed into the transducer well. The transducer head was held in place in the well with a clamp attached to a ring stand (Fig. 2). In an effort to decrease testing error, the same researcher positioned the US head in the transducer well for all machines tested. Four intensity settings (0.5, 1.0, 1.5, and 2.0 W/[cm.sup.2]) were tested on each US unit using a continuous waveform The shape of a signal. See wavelength, sine wave and square wave. . A complete list of manufacturers, models, and numbers of machines tested is presented in Table 1. Machines tested had one of the following soundhead's effective radiation area sizes: 4, 5, 6, 8.5, or 10 [cm.sup.2].

[FIGURE 2 OMITTED]

Calculations

The difference between the power output registered on the UW-2 (measured power output) and the intensity output indicated on the US unit (indicated power See Indicated horse power, under Horse power.

See also: Indicated
 output) was expressed as a percentage of error using the formula: [ (measured power output - indicated intensity output) /indicated intensity output] x 100. This percentage of error was calculated at each intensity setting and recorded for each US unit tested.

Each machine's timer was tested for accuracy after the output measurements were taken. Timers were tested at 5- and 10-minute intervals after placing the transducer heads into beakers of water. The difference between the time registered on the US unit (indicated time) and the actual time recorded on the stopwatch (measured time) was recorded for every US unit tested using the formula: [(measured time - indicated time) /indicated time] x 100.

Results

Power Output

The results showed that 32 (39%) of the 83 machines were not within the [+ or -] 20% standard of calibration at one or more of the tested settings. Of these machines, 15 (18%) were above the [+ or -] 20% standard and 17 (21%) were below the -20% standard for at least one output setting. Of these 32 machines, 26 (81%) were outside the standard at 2 or more settings. Three (9%) of the 32 machines had no output at any setting. Of the 51 machines that were considered to be inside the calibration standard, 8 (16%) were exactly 20% away from the expected output for at least one setting. Table 2 lists the brand names of the US machines tested and the numbers of machines within and outside the standard.

The correlation between the percentage of error for each US machine versus the age of each machine was examined. A statistically significant correlation was found between percentage of error for each US machine and age of each machine only for an intensity of 2.0 W/[cm.sup.2] (r=.29, P=.01). No statistically significant correlation was found for intensities of 0.5, 1.0, and 1.5 W/[cm.sup.2]. The correlations between how often each US machine was checked and how often it was used versus the percentage of error for each machine also were examined. No statistically significant correlation was found.

No difference was found when the 5 major brand names of the US machines tested were compared with the number of machines outside the standard using a chi-square test chi-square test: see statistics.  with a Yates correction. However, as indicated in Figure 3, some brands had a greater number of machines within the standard than other brands.

Timers

Of the 83 US machines tested, 25 (30%) had mechanical timers and 58 (70%) had digital timers. All of the digital timers were within the [+ or -] 10% calibration standard. Seven (28%) of the 25 mechanical timers were outside the standard at the 5-minute interval, and 6 (24%) were outside the standard at the 10-minute interval (Fig. 4).

Survey

The most commonly used intensities were reported as 1.1 to 1.5 W/[cm.sup.2] (75%). The ages of the US machines ranged from 1 to 20 years, with a mean age of 8.1 years (SD=5.5). Table 2 shows the mean age of the US machines for each brand. The frequency of use for the US machines was as follows: 4.1 to 5.0 hours per week=6%, 3.1 to 4.0 hours per week=8%, 1.1 to 2.0 hours per week=17%, 0 to 1.0 hour per week=22%, greater than 5.0 hours per week=23%, and 2.1 to 3.0 hours per week=24%. All 83 machines were reported to be "checked" at least annually. Clinicians surveyed reported that they "did not know" the variables (ie, power output, electrical supply and safety, and timer accuracy) tested on 38% of the US machines. Power output was reported to be "checked" on 46% of the US machines, and electrical supply and safety were the only variables reported to be "checked" on 14% of the machines.

Discussion

The clinicians assumed that technicians, either in-house personnel or outsource vendors, were conducting periodic "checks" on the US machines to ensure proper power output, electrical safety, and timer accuracy. Actually, all 3 variables, just one variable, or a combination of variables might have been "checked."

We found that 32 (39%) of the tested US machines had variables outside the standard for at least one setting. Of these machines, 15 (18%) were above the +20% standard and 17 (21%) were below the -20% standard for at least one output setting. The 17 US machines below the -20% standard may only be denying the patient effective treatment. However, the 15 US machines above the +20% standard could be a potential harm to the patient. An additional 8 (10%) of the US machines were within the [+ or -] 20% standard but on the borderline borderline /bor·der·line/ (-lin) of a phenomenon, straddling the dividing line between two categories.
borderline 
. These results indicate that the percentage of US machines outside the standard has decreased over the last 3 decades when compared with results previously published. This finding could be due to a greater awareness by clinicians of the importance of periodic "checks." No tests were run to identify the source of the problem for the US machines outside the standard. These results indicate that clinicians should request in-house technicians (eg, biomedical bi·o·med·i·cal
adj.
1. Of or relating to biomedicine.

2. Of, relating to, or involving biological, medical, and physical sciences.
 services) or outsource vendors to check the US machine annually for correct power output. If the US machine is found not to be within the [+ or -] 20% standard, the machine should be calibrated by the technician or sent back to the manufacturer for calibration. Whether this matters in terms of clinical effectiveness is not known.

With the exception of one weak correlation coefficient between the age of the machines and the percentage of error at 2 W/[cm.sup.2] (r=.29, P=.01), there was no statistically significant association among the amount of time the machines were used per week, their age, and the percentage of error of power output. This finding indicates that age and usage of US machines are not good predictors of the machines' correct output. Results also indicated no significant correlation between how often the machines were "checked" and the percentage of error of power output. Because all machines tested were reported to be "checked" at least annually, an assumption cannot be made in this study as to whether frequency of testing has an effect on power output being within the standard.

A relationship regarding the output accuracy of certain brands of US machines was also observed. In our study, Enraf-Nonius ([dagger]) and Dynatronics ([double dagger double dagger
n.
A reference mark () used in printing and writing. Also called diesis.

Noun 1.
]) brands exhibited a greater percentage of machines that were within the standard (n=25; >85%) than Chattanooga, ([section]) Rich-Mar, ([parallel]) and Mettler ([#]) brands (n=51; <50%) (Fig. 3).

All digital timers were within the standard at both time intervals tested, whereas approximately one fourth of the mechanical timers were outside the standard (Fig. 4). This finding suggests that machines with digital timers are preferable to ensure correct treatment duration.

Our survey revealed that a large number of clinicians were unaware of the variables checked for each machine. Because clinicians are responsible for safe delivery of US dosage to the patient, we believe they need to be more aware of the variables that are to be tested to ensure safe delivery of US. Therapists and others, in our view, could benefit from systematic training in their education curricula on specific variables of US machines (eg, power output) in need of periodic checks. Continuing education continuing education: see adult education.
continuing education
 or adult education

Any form of learning provided for adults. In the U.S. the University of Wisconsin was the first academic institution to offer such programs (1904).
 programs need to be available to educate clinicians on this topic.

We suggest that further studies in US calibration should involve testing US units at different frequencies (1 MHz (MegaHertZ) One million cycles per second. It is used to measure the transmission speed of electronic devices, including channels, buses and the computer's internal clock. A one-megahertz clock (1 MHz) means some number of bits (16, 32, 64, etc.  versus 3 MHz), because frequency is important in regard to the depth of the tissue targeted. Evaluating the differences in machine accuracy when using pulsed versus continuous waveforms could also be investigated. Calibration studies could also be conducted throughout the country to determine whether there is a consistent trend in accuracy of calibration in different regions.

Conclusion

We believe that clinicians should be aware that the intensity displayed on US units is not always a direct indication of the actual output being emitted. In addition, although most new US units have digital timers, we strongly recommend when purchasing used equipment that the machines have digital timers.

According to according to
prep.
1. As stated or indicated by; on the authority of: according to historians.

2. In keeping with: according to instructions.

3.
 our data, improvement has been made regarding the accuracy of therapeutic US when compared with past research findings. This improvement may be due to several factors. One factor may have been an increase in awareness by US technicians and manufacturers of the importance of machine accuracy. The differences may also have been attributable to testing machines testing machine

Machine used in materials science to determine the properties of a material. Machines have been devised to measure tensile strength, strength in compression, shear, and bending (see strength of materials), ductility, hardness, impact strength (
 in a different region than previous studies. However, more than one third of machines tested in this study were outside the standard for power output, as were almost one fourth of mechanical timers. Thus, further improvement in the accuracy of US machine calibration is needed. Clinicians should not hesitate to request that US machines be checked for power output and timer accuracy, which would benefit the safety of the patient and decrease the liability of the institution. Proper calibration would help ensure that patients receive a more accurate US dosage and, therefore, safe and appropriate treatment.
Figure 3.

Comparison of 5 major ultrasound machines (n > 8) tested versus the
number of ultrasound machines within and outside of the standard.
See Table 1 footnotes for manufacturers' addresses.

                           No. of Machines

Brand Name      Within Standard    Outside Standard

Chattanooga            17                 15

Enraf-Nonius           14                  1

Rich-Mar                5                  6

Dynatronics             8                  2

Mettler                 3                  5

Note: Table made from bar graph
Figure 4.

Comparison of mechanical versus digital timers within and outside
the standard.

                                 No. of Machines

Type of Timer         Within Standard    Outside Standard

Mechanical 5 min.            18                  7

Digital 5 min.               58                  0

Mechanical 10 min.           19                  6

Digital 10 min.              58                  0

Note: Table made from bar graph
Table 1.
Details of Ultrasound Machines Assessed in the Study

                                                    No. of
                                                    Machines
Manufacturer              Model                     Tested

Amrex (a)                 Synchrosonic 54            2

Bosch (b)                 Sonomed 4                  1

Chattanooga (c)           Forte 200 Combo            5
                          Forte 400 Combo            4
                          Intellect Legend Combo     5
                          Intellect 700             11
                          Intellect 240              1
                          Intellect 225P             1
                          Forte CB                   1
                          Intellect 200              1
                          Vectra                     1
                          Forte US                   1
                          Intellect US               1

Dynatronics (d)           Dynatron 950               9
                          Dynatron 800               1

Enraf-Nonius (e)          Sonopuls 434               8
                          Sonopuls 464               4
                          Sonopuls 590               3

Excel (f)                 Ultra IV                   1

Linquist (g)              Chronosonic Ultrasound     1

Mettler (h)               Sonicator 706              4
                          Sonicator II               2
                          Sonicator 730              1
                          Sonicator 720              1

Mid-Canada Medical (i)    Medisound                  1

PTI (j)                   Omnisound 2500C            1

Rich-Mar (k)              Rich-Mar VI                2
                          Rich-Mar X                 2
                          3P                         2
                          Thera-touch                1
                          Rich-Mar V                 1
                          Rich-Mar 2060              1
                          510                        1
                          HV-II                      1

(a) Amrex Electrotherapy Equipment, 641 E Walnut St, Carson, CA
90746.

(b) Bosch Corp, 2800 S 25th Ave, Broadview, IL 60155.

(c) Chattanooga Group Inc, 4717 Adams Rd, Hixson, TN 37343.

(d) Dynatronics Corp, 7030 Park Centre Dr, Salt Lake City, UT
84121.

(e) Enraf-Nonius BV, Rontgenweg 1/PO Box 810, 2600 AV Delft, the
Netherlands.

(f) Excel Tech Ltd, 2892 Portland Dr, Oakville, Ontario, Canada
L6H 5W8.

(g) Linquist, address unavailable.

(h) Mettler Electronics Corp, 1333 S Claudina St, Anaheim, CA
92805.

(i) Mid-Canada Medical, 6200A Tomken Rd, Mississauga, Ontario,
Canada L5T 1X7.

(j) Power Technologies Inc, 1482 Erie Blvd, PO Box 1058,
Schenectady, NY 12301-1058.

(k) Rich-Mar Corp, PO Box 879, Inola, OK 74036.
Table 2.
Number of Ultrasound Machines Outside the Standard at Only One
Setting and at Two or More Settings and the Number of Machines
Within the Standard for Each Brand Name Tested

                      Total No.      Mean Age    No. Out at
Brand Name (a)        of Machines    (y)         Only One Setting

Chattanooga           32              7.3
Enraf-Nonius          15             11.3        1
Rich-Mar              11              8.4        2
Dynatronics           10              3.4
Mettler                8             10.8        2
Amrex                  2              5.5        1
Bosch                  1             14.0
Excel                  1              2.0
Lindquist              1             20.0
Mid-Canada Medical     1              5.0
PTI                    1              6.0
Total                 83              8.1        6

                      No. Out at Two      No. Within
Brand Name (a)        or More Settings    Standard

Chattanooga           15                  17
Enraf-Nonius                              14
Rich-Mar               4                   5
Dynatronics            2                   8
Mettler                3                   3
Amrex                                      1
Bosch                  1
Excel                                      1
Lindquist              1
Mid-Canada Medical                         1
PTI                                        1
Total                 26                  51

(a) See Table 1 footnotes for manufacturers' addresses.


* Bio-Tek Instruments Inc. Highland Park Highland Park.

1 City (1990 pop. 30,575), Lake co., NE Ill., a suburb of Chicago on Lake Michigan; inc. 1869. It is a retail business and medical center for the North Shore area.
, Box 998, Winooski, VT 05404.

([dagger]) Enraf-Nonius BV, Rontgenweg 1/PO Box 810, 2600 AV Delft Delft (dĕlft), city (1994 pop. 91,941), South Holland prov., W Netherlands. It has varied industries and is noted for its ceramics (china, tiles, and pottery) known as delftware. Founded in the 11th cent. , the Netherlands.

([double dagger]) Dynatronics Corp, 7030 Park Centre Dr, Salt Lake City, UT 84121.

([section]) Chattanooga Group Inc, 4717 Adams Rd, Hixson, TN 37343.

([parallel]) Rich-Mar Corp, PO Box 879, Inola, OK 74036.

(#) Mettler Electronics Corp, 1333 S Claudina St, Anaheim, CA 92805.

References

(1) Stewart HF, Abzug JL, Harris GR. Considerations in ultrasound therapy ultrasound therapy Mainstream medicine The application of ultrasound waves to soft tissue to heat and relax injured tissue and disperse edema  and equipment performance. Phys Ther. 1980;60:424-428.

(2) ter Haar G. Basic physics of therapeutic ultrasound. Physiotherapy physiotherapy: see physical therapy. . 1978;64:100-103.

(3) Mortimer AJ, Dyson M. The effect of therapeutic ultrasound on calcium uptake in fibroblasts Fibroblasts
A type of cell found in connective tissue; produces collagen.

Mentioned in: Skin Grafting
. US Med Biol. 1988;14:499-506.

(4) Lehmann JF. Clinical evaluation clinical evaluation Medtalk An evaluation of whether a Pt has symptoms of a disease, is responding to treatment, or is having adverse reactions to therapy  of a new approach in the treatment of contracture contracture /con·trac·ture/ (-cher) abnormal shortening of muscle tissue, rendering the muscle highly resistant to passive stretching.  associated with hip fracture hip fracture Orthopedic surgery A femoral fracture which affects 1/6 white ♀–US during life Epidemiology 250,000/yr–US Specifics Proximal femur; 90+% femoral neck, intertrochanteric; 5-10% are subtrochanteric Risk factors Tall, thin ♀,  after internal fixation internal fixation
n.
The stabilization of fractured bony parts by direct fixation to one another with surgical wires, screws, pins, or plates.
. Arch Phys Med Rehabil. 1961;42:95.

(5) Wessling KC, DeVane DA, Hylton CR. Effects of static stretch versus static stretch and ultrasound combined on triceps surae The triceps surae is a term given by some anatomists to the gastrocnemius and soleus muscles together as they both insert into the calcaneus, the bone of the heel of the human foot, and form the major part of the muscle of the back part of the lower leg (the calf; otherwise known  muscle extensibility in healthy women. Phys Ther. 1987;67:674-679.

(6) Harvey W, Dyson M, Pond JB, Grahame R. The stimulation of protein synthesis Protein synthesis is the creation of proteins using DNA and RNA. Biological and artificial methods for creation of proteins differ significantly.
  • For biological protein synthesis, see protein biosynthesis.
  • For artificial protein synthesis, see peptide synthesis.
 in human fibroblasts by therapeutic ultrasound. Rheumatol Rehabil. 1975;14:237.

(7) Middlemast S, Chatterjee DS. Comparison of ultrasound and thermotherapy ther·mo·ther·a·py
n.
Medical therapy involving the application of heat.


thermotherapy
 for soft tissue injuries Soft tissue injury is damage of the soft tissue of the body. These types of injuries are a major source of pain and disability. The four fundamental tissues that are affected are the epithelial, muscular, nervous and connective tissues. . Physiotherapy. 1978;64:331-332.

(8) Frieder SJ, Weisberg B, Fleming B, et al. A pilot study: the therapeutic effect of ultrasound following partial rupture rupture, in medicine: see hernia.  of Achilles tendon Achilles tendon
n.
The large tendon connecting the heel bone to the calf muscle of the leg. Also called calcanean tendon, heel tendon.
 in male rats. J Orthop Sports Phys Ther. 1988;10:39-46.

(9) Enwemeka CS. The effects of therapeutic ultrasound on tendon healing: a biomechanical Biomechanical may refer to:
  • Bioengineering
  • Biomaterial
  • Biomechanical (band)
  • Biomechanics
  • Biomechanoid
  • Biorobotics
  • Bioship
  • Cyborg
  • Organic (model)
 study. Am J Phys Med Rehabil. 1989;6:283-287.

(10) Enwemeka CS, Rodriguez O, Mendosa S. The biomechanical effects of low intensity ultrasound on healing tendons. Ultrasound Med Biol. 1990;16:801-807.

(11) Jackson BA, Schwane JA, Starcher Starch´er   

n. 1. One who starches.
 BC. Effect of ultrasound therapy on the repair of Achilles tendon injuries in rats. Med Sci Sports Exerc. 1991;23:171-176.

(12) Draper DO, Castel JC, Castel D. Rate of temperature increase in human muscle during 1 Hz and 3 MHz continuous ultrasound. J Orthop Sports Phys Ther. 1995;22:142-150.

(13) Deforest de·for·est  
tr.v. de·for·est·ed, de·for·est·ing, de·for·ests
To cut down and clear away the trees or forests from.



de·for
 RE. Effects of ultrasound on growing bone: an experimental study. Arch Phys Med Rehabil. 1953;34:21.

(14) Payton OD, Lamb RL, Kasey ME. Effects of therapeutic ultrasound on bone marrow in dogs. Phys Ther. 1975;55:20-27.

(15) Kitchen SS, Partridge partridge, common name applied to various henlike birds of several families. The true partridges of the Old World are members of the pheasant family (Phasianidae); the common European or Hungarian species has been successfully introduced in parts of North America.  CJ. A review of therapeutic ultrasound: background and physiological effects. Physiotherapy. 1990;76:593-595.

(16) Dyson M. Mechanisms involved in therapeutic ultrasound. Physiotherapy. 1987;73:116-119.

(17) US Department of Health, Education, and Welfare, Food and Drug Administration. Ultrasonic therapy products: radiation safety performance standard. US Federal Register. Title 21, part 1050. 1978;43: 7166-7172.

(18) Health Protection Branch, Health and Welfare Canada Health and Welfare Canada is a former Canadian federal department established in 1944 and split into two separate departments, Health Canada and Human Resources and Labour Canada, in June 1993 by Prime Minister Kim Campbell. . Radiation emitting device regulations (Amendment PC 1981-908). Canada Gazette The Canada Gazette is an official publication by the government of Canada that publishes all laws and Orders-in-Council issued by the government. It also contains other information on things such as hearing and tribunals, proposed changes and any thing else the government . April 1981, Part II;115(8):1121-1126.

(19) Particular Requirements for the Safety of Ultrasonic Therapy Equipment. Geneva Geneva, canton and city, Switzerland
Geneva (jənē`və), Fr. Genève, canton (1990 pop. 373,019), 109 sq mi (282 sq km), SW Switzerland, surrounding the southwest tip of the Lake of Geneva.
, Switzerland: International Electrotechnical Commission; 1984. International Electrotechnical Commission Publication 601-2-5.

(20) Stewart HF, Harris GR, Herman BA, et al. Survey of use and performance of ultrasonic therapy equipment in Pinellas County, Florida Pinellas County is a county located in the state of Florida. Its county seat is Clearwater, Florida6, and its largest city is St. Petersburg. The county is contained entirely within area code (727), except for sections of Oldsmar, which has area code 813. . Phys Ther. 1974;54:707-715.

(21) Snow CJ. Ultrasound therapy traits in Manitoba and Northwestern Ontario Northwestern Ontario is the region within the Canadian province of Ontario which lies north and west of Lake Superior, and west of Hudson Bay and James Bay. It includes most of subarctic Ontario. : performance evaluation Performance evaluation

The assessment of a manager's results, which involves, first, determining whether the money manager added value by outperforming the established benchmark (performance measurement) and, second, determining how the money manager achieved the calculated return
. Physiotherapy Canada. 1982;34: 185-189.

(22) Lloyd JJ, Evans JA. A calibration survey of physiotherapy ultrasound equipment in North Wales North Wales (known in some archaic texts as Northgalis) is the northernmost unofficial region of Wales, bordered to the south by Mid Wales and to the east by England. . Physiotherapy. 1988;74:56-61.

(23) Pye SD, Milford C. The performance of ultrasound physiotherapy machines in Lothian Region Noun 1. Lothian Region - a district in southeast central Scotland (south side of the Firth of Forth) and the location of Edinburgh
Scotland - one of the four countries that make up the United Kingdom of Great Britain and Northern Ireland; located on the northern
, Scotland, 1992. Ultrasound Med Biol. 1994;20:347-359.

(24) Guirro R, Serrao F, Elias D, Bucalon AJ. Calibration of therapeutic ultrasound equipment. Physiotherapy. 1997;83:419-422.

This study was approved by the Amarillo Institutional Review Board of Texas Tech University Health Sciences Center The Texas Tech University Health Sciences Center offers Schools of Allied Health Sciences, Biomedical Sciences, Medicine, Nursing, and Pharmacy. The HSC has campuses located in Lubbock, as well as in Abilene, Amarillo, El Paso, and Odessa. .

This article was submitted March 12, 2001, and was accepted July 24, 2001.

PA Artho, PT, MPT MPT Maryland Public Television
MPT Modern Portfolio Theory (investing)
MPT Ministry of Posts and Telecommunications
MPT Message-Passing Toolkit
MPT Master of Physical Therapy
MPT Mitochondrial Permeability Transition
; JG Thyne, PT, MPT; BP Warring, PT, MPT; and CD Willis, PT, MPT, were students in the Master of Physical Therapy The Master of Physical Therapy (MPT) is a postbaccalaureate degree conferred upon successful completion of an accredited Physical therapy professional education program. Successful candidates are then qualified to apply for and take the Physical Therapy national licensure exam (in  Program, Texas Tech University Health Sciences Center, Amarillo, Tex, at the time of the study.

JM Brismee, PT, MS, OCS OCS - Object Compatibility Standard , is Assistant Professor, School of Allied Health, Texas Tech University Health Sciences Center, 1400 Wallace Blvd, Amarillo, TX 79106 (USA) (brismee@cortex.ama.ttuhsc.edu), and Faculty Member, International Academy of Orthopaedic Medicine (IAOM-US), Tucson, Ariz. Address all correspondence to Mr Brismee at the first address.

NS Latman, PhD, is Associate Professor, School of Allied Health, Texas Tech University Health Sciences Center.

Mr Brismee provided concept/research design, Mr Warring and Mr Willis provided writing. Mr Artho and Mr Thyne provided data collection, Mr Thyne provided data analysis and facilities/equipment. Mr Artho and Mr Willis provided project management. Mr Warring provided fund procurement, and Mr Artho and Mr Warring provided clerical support. Dr Latman provided consultation (including review of manuscript before submission). The authors thank Lanny Traves from Baptist St Anthony's Hospital, Amarillo, Tex, for the use of the Bio-Tek Digital Ultrasound Wattmeter (model UW-2). They also thank Steve Sawyer, PT, PhD, Assistant Professor, Texas Tech University Health Sciences Center, Lubbock, Tex, for assistance with statistical programs and problem solving problem solving

Process involved in finding a solution to a problem. Many animals routinely solve problems of locomotion, food finding, and shelter through trial and error.
.
COPYRIGHT 2002 American Physical Therapy Association, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2002, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Author:Latman, Neal S
Publication:Physical Therapy
Article Type:Statistical Data Included
Geographic Code:1USA
Date:Mar 1, 2002
Words:4098
Previous Article:Relationships among lifting ability, grip force, and return to work. (Research Report).
Next Article:Fear of falling. (Update).
Topics:



Related Articles
Effects of ultrasound and trolamine salicylate phonophoresis on delayed-onset muscle soreness. (includes commentary and author response)
In vitro effects of therapeutic ultrasound on the nucleus of human fibroblasts.
Phonophoretic delivery of 10% hydrocortisone through the epidermis of humans as determined by serum cortisol concentrations. (includes commentary and...
Effects of energy-matched pulsed and continuous ultrasound on tumor growth in mice.
The use of therapeutic ultrasound by physical therapists in Dutch primary health care.
A Review of Therapeutic Ultrasound: Effectiveness Studies.
A Review of Therapeutic Ultrasound: Biophysical Effects.
Philadelphia panel evidence-based clinical practice guidelines on selected rehabilitation interventions for neck pain. (Specail Issue).
Philadelphia panel evidence-based clinical practice guidelines on selected rehabilitation interventions for shoulder pain. (Special Issue).
Ultrasound produced by a conventional therapeutic ultrasound unit accelerates fracture repair.(Research Report)(Clinical report)

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles